ALLSHENG ALLSHENG
MD200-2
  • MD200-2
Sample Concentration — MD200-2

MD200-2 sample concertration is mainly used for the concentration and preparation of large batches of samples, such as sample preparation in drug screening, hormone analysis, liquid phase, gas phase and mass spectrometry analysis. Working principle: By blowing nitrogen gas into the surface of the heated sample, the solvent in the sample is quickly evaporated and separated, so as to achieve the purpose of anaerobic concentration of the sample and keep the sample more pure.

  • ALLSHENG ALLSHENG
    Features
  • ALLSHENG ALLSHENG
    Technical parameter
  • ALLSHENG ALLSHENG
    Order information

•High precision and wide range of temperature control

•Blowing a large number of samples at one time

•Blowing each sample independently

•Amount of airflow can be adjusted

•Easy operation; safe and reliable to use

•Low nitrogen consumption of 330 ml/min/sample

•The heater enables the sample to be quickly and efficiently heated to the evaporation temperature, and at the same time the gas is blown from the gas chamber to the surface of the solution through the gas needle, which promotes rapid evaporation of the solution and concentration of the sample

•The position of the gas needle in the gas chamber can be changed to make it suitable for different test tubes. The standard gas needle length is 150 mm

•The height of the air cavity can be adjusted, this product has a unique feature: the liquid level position of the concentrated sample can be observed at any time when the sample is concentrated

•When concentrating toxic solvents, the entire system can be placed in a fume hood

•Automatic fault detection and alarm function

•Standard with air chamber and special adjustable stand

•Built-in over temperature protection device

•Real time temperature display, time decrement display stand

  • Model
    MD200-2
  • Temperature control range
    RT.+5 ℃~150 ℃
  • Temp. control accuracy @40~100 °C
    ±0.5 ℃
  • Temp. control accuracy @100~150 °C
    ±1 ℃
  • Temp. accuracy for blocks @40 °C
    ±0.3 ℃
  • Display accuracy
    0.1 ℃
  • Heating time (40~150 °C)
    ≤30 min
  • Standard blocks
    2
  • Needle length
    150 mm
  • Nitrogen consumption
    0~15 L/min
  • Dimension (mm)
    280×240×500 mm
  • Code
    Description
  • AS-06070-00
    MD200-2, AC220 V/AC120 V, 50/60 Hz, 400 W