Model
Feyond-M300/M380/M400/M480/M500/M580
-
ABS
-
Light source
Xenon lamp
-
Wavelength accuracy
2 nm
-
Wavelength repeatability (SD)
0.2 nm
-
Half width (FWHM)
<2.5 nm
-
Detector
PD
-
Wavelength range
190-1000 nm, 1 nm step
-
Measure range
0 - 4 OD
-
Resolution
0.0001 OD
-
Accuracy @450 nm
96-precision mode: ±(1.0%+0.003 Abs) @ (0.0-2.0 Abs] ±2.0% @ (2.0-3.0 Abs]
-
Repeatability @450 nm
CV < 1.0% fast (0.0 - 3.0]; CV < 0.5% accurate (0.0 - 3.0]
-
Stray light
0.1% @220 nm
-
Linear @450 nm
R^2 ≥ 0.999 @ [0.0 - 3.0 Abs
-
Reading time
96-well plate: fast <15 s (A1 to H1)
-
FL
-
Reading mode
Top reading
-
Excitation light source
Xenon lamp
-
Detector
PMT
-
Wavelength range
EX: 200 - 1000 nm; EM: 270 - 850 nm
-
Filter EX/EM
3 groups: EX470/EM525, EX523/EM564, EX624/EM692 (other wavelengths can be replaced)
-
Linear dynamic range
6 logs
-
Detection limit
2 pM (optimization condition)
-
LUM
-
Detector
PMT
-
Detection limit
≤25 amol/well @96-well white plate flash ATP
-
Linear dynamic range
6 logs
-
Crosstalk
≤ 0.005% (ATP flash @96-well white plate)
-
Wavelength range
200-850 nm
-
TRF
-
Filter EX/EM
1 group: EX365/EM612
-
Detection limit
0.02pM (optimization condition)
-
FP
-
Filter EX/EM
1 group: EX485/EM530
-
Detection limit
sd≤5 mP@1 nm fluorescein sodium